Category: Laboratory and Electrochemistry (WQL)
The 4310W probe can only be calibrated with a 1,413 uS/cm conductivity standard (060906) (Select in ''Option'' drop down menu).
Wireless MultiLab probes do not include a WA-S wireless sensor module.
One WA-S wireless sensor module must be installed at the top of each wireless sensor, while one WA-M wireless meter module is needed per MultiLab meter. We suggest the MultiLab WA Kit if wireless sensors will be used, as it includes one WA-S wireless sensor module, one WA-M wireless meter module, one charger for WA-S modules, and one USB power supply for the WA-S charger.
Wireless sensors can only be used with the MultiLab 4010-1W, MultiLab 4010-2W, and the MultiLab 4010-3W. Wireless sensors can be used with the MultiLab 4010-1 if the latest instrument firmware has been installed - see the Software and Firmware Downloads page.
Please see the MultiLab Ordering Guide if ordering a complete system, or contact us if you have any questions.
Cell Constant |
0.475 cm-1 ±1.5 % |
---|---|
Equipment used with | MultiLab Instruments |
Measurement Range |
Conductivity: 10 µS/cm to 2,000 mS/cm | Resistivity: 0.5 Ohm cm to 100 kOhm cm | Salinity: 0.0 to 70.0 ppt | Total Dissolved Solids: 0 to 1,999 mg/L; 0.0 to 199.9 g/L | Temperature: 0 to 100ºC |
Minimum Immersion Depth | 36 mm |
Parameters Measured |
Conductivity, Temperature |
Storage Temperature | 0 to 50 °C |
Warranty |
2 years |